Connect the transducer – Rockwell Automation 1440-SDM02-01RA XM-124 Standard Dynamic Measurement Module Manuel d'utilisation
Page 34

34
Rockwell Automation Publication 1440-UM001C-EN-P - May 2014
Chapter 1
Install the XM-124 Standard Dynamic Measurement Module
Table 4 - Configuring Buffered Output Input Range
Connect the Transducer
The XM-124 module can accept inputs from any Allen-Bradley non-contact
eddy current probe, a standard IEPE accelerometer, a velocity transducer, AC
voltage output, or a DC voltage output measurement device.
Connect an IEPE Accelerometer
The following figures show the wiring of an IEPE accelerometer to the terminal
base unit.
IMPORTANT
The voltage operating range of the buffered outputs must be configured to
coincide with the corresponding transducer bias range. This operating
range is configured by placing a jumper from terminal 5 (channel 1) and
terminal 22 (channel) to either terminal 6 (Positive Buffer Bias) or terminal
21 (Buffer -), depending on the transducer. See
. The buffered
output operating range is configured independently per channel.
Transducer
Input Range
Channel
Connect Terminal
To Terminal
Negative Bias
-24…9V
1
5
21
2
22
21
Positive Bias
-5…24V
1
5
6
2
22
6
Non-Bias
-5…9V
1
----
----
2
----
----
ATTENTION: You may ground the cable shield at either end of the cable.
Do not ground the shield at both ends. Recommended practice is to
ground the cable shield at the terminal base and not at the transducer.
Any convenient chassis terminal may be used (see
).
IMPORTANT
The internal transducer power supply is providing power to the IEPE
accelerometer. Make certain the IEPE Power parameter is enabled. Refer
to
Transducer Object (Class ID 328H) on page 154
.
IMPORTANT
A jumper from terminal 5 to terminal 6 is required for channel 1 buffered
output. A jumper from terminal 22 to terminal 6 is required for channel 2
buffered output. See