Metrohm 730 Sample Changer Manuel d'utilisation
Page 94

4.3 Menu Organization
730 Sample Changer, Detailed Description
91
Parameter
>dosing unit def.
Submenu for dosing unit settings
Open the submenu with <ENTER>
>dosing unit def.
dosing unit
1
Select dosing unit
1…12
After entering the address of the dosing unit (see page 23) and con-
firming with <ENTER> it will be displayed in the first menu line.
>dosing unit
1
dos.rate:
160 ml/min
Set the dosing speed
0.01…160 ml/min,
max.
>dosing unit
1
fill.rate:
160 ml/min
Set the filling speed
0.01…160 ml/min,
max.
The following entries are only valid for 700 Dosinos. For details
about Dosinos and dosing units, see page 109ff.
>dosing unit
1
dosing
port 1
Define dosing outlet
1…4
>dosing unit
1
filling
port 2
Define filling inlet
1…2…4
>dosing unit
1
rinsing
port 2
Define rinsing inlet (when chang-
ing dosing units)
1…2…4
>dosing unit
1
preparation
port 1
Define outlet for the preparation
cycle
1…4
>dosing unit
1
drain
port 4
Define air inlet for emptying
1…4