Metrohm 730 Sample Changer Manuel d'utilisation
Page 111
Advertising

4.5
Sample Racks
730 Sample Changer, Detailed Description
108
Sample method for the use of special beakers:
Electrode calibration in a start sequence
LIFT: 1 :
shift mm
MOVE 1 :
spec.1
Spec.1 = rinsing beaker
LIFT: 1 :
rinse mm
PUMP 1.1 :
3 s
rinse electrode
WAIT
2 s
LIFT: 1 :
shift mm
MOVE 1 :
spec.2
spec.2 = buffer 1
LIFT: 1 :
work mm
STIR: 1 :
on s
CTL:Rm: START device1
start measuring buffer 1
SCN:Rm:
endmeter 1
measurement finished? then ...
STIR: 1 :
off s
LIFT: 1 :
shift mm
MOVE 1 :
spec.1
spec.1 = rinsing beaker
LIFT: 1 :
rinse mm
PUMP 1.1 :
3 s
rinse electrode
WAIT
2 s
LIFT: 1 :
shift mm
MOVE 1 :
spec.3
spec.3 = buffer 2
LIFT: 1 :
work mm
STIR: 1 :
on s
CTL:Rm: START device1
start measuring buffer 2
SCN:Rm:
endmet 1
measurement finished? then...
STIR: 1 :
off s
...