Metrohm 730 Sample Changer Manuel d'utilisation
Page 149

4.10 Remote Control Commands
730 Sample Changer, Detailed Description
146
4.10.2.30
I
nfo.ActualInfo.Pump.1.State
etc. up to .4
Current pump state (on/off).
Pump 1 and 2 on tower 1, Pump 3 and 4 on tower 2.
4.10.2.31
I
nfo.ActualInfo.Buret.1.State
read only
I
nfo.ActualInfo.Buret.1.Position
read only
I
nfo.ActualInfo.Buret.1.Cock
read only
I
nfo.ActualInfo.Buret.1.Type
read only
I
nfo.ActualInfo.Buret.1.Volume
read only
etc. until .12
Current data of dosing drive.
State:
Status (ready/busy)
Position:
Piston position in mL
Cock:
Cock position
Type:
Type of dosing unit (685/700)
Volume:
Burette volume
4.10.2.32
I
nfo.ActualInfo.Inputs.Status
read only
Status of the Input lines (Input0…7) of the Remote interface.
$Q sends the signal state as decimal number e.g.
10
⇒
00001010 binary
⇒
2
1
+ 2
3
⇒
Input1 and Input3 active
(active = low, inactive = high)
See also page 119ff.
4.10.2.33
I
nfo.ActualInfo.Outputs.Status
read only
Status of the output lines (Output0…13) of Remote interface. See
4.10.2.32.
4.10.2.34
I
nfo.ActualInfo.Display.L1
read only
I
nfo.ActualInfo.Display.L2
read only
Text of the first and second line of the LCD display.
4.10.2.35
I
nfo.ActualInfo.Counter.Sample
read only
I
nfo.ActualInfo.Counter.Maximum
read only
Currently processed sample number and total number of samples.