5.4 Diagnosis
730 Sample Changer, Appendix
173
Configuration
Identification in the Display
Tower 1
address 0x86 type 3
Tower 2
address 0x87 type 3
Dos 1 - 4
(729 / Address 1)
address 0x91 type 3
address 0x92 type 3
Dos 5 - 8
(729 / Address 2)
address 0xA1 type 3
address 0xA2 type 3
Dos 9 - 12 (729 / Address 3)
address 0xB1 type 3
address 0xB2 type 3
>external bus test
address 0x86 type 3
•
Press <ENTER> several times and compare display with the
list above until
diagnosis
>beeper test
5
5.4
.4.9
.9 Beeper
•
Prepare instrument for diagnosis (see Chap. 5.4.2).
•
If necessary, press <ê> several times until
diagnosis
>beeper test
•
<ENTER>
The beeper is turned on and off in an endless loop.
>beeper test
•
The test can be exited by pressing the <QUIT> or <STOP>
keys.
diagnosis
>rack code test
5.4.10 Rack Code Recognition
Rack Code Recognition
This diagnosis step checks the functionality of the sensors with re-
gard to their ability to automatically recognize the attached rack.
•
Note the code (magnets) on the underside of all the racks being
used. The following diagram shows the magnet fixture on the
underside of a rack.