4.9 Operation via RS232 Interface
730 Sample Changer, Detailed Description
129
Program-specific Messages:
E28
Wrong object call
E29
Wrong value or no value allowed
E30
Wrong trigger
RS Receive Errors:
E36
Parity error
Exit: <QUIT> and set the same parity for both instruments.
E37
Stop bit error
Exit: <QUIT>and set the same stop bit for both instruments.
E38
Overrun error. At least 1 character could not be read.
Exit: <QUIT>.
E39
Internal receive buffer full (>82 characters).
Exit: <QUIT>.
RS Send Errors:
E40
DSR=OFF. No proper handshake for more than 1s. Exit: <QUIT>.
Is the receiver switched on and ready to receive?
E41
DCD=ON. No proper handshake for more than 1s. Exit: <QUIT>.
Is the receiver switched on and ready to receive?
E42
CTS=OFF. No proper handshake for more than 1s. Exit: <QUIT>.
Is the receiver switched on and ready to receive?
E43
Transmission has been interrupted for at least 3 s with XOFF.
Exit: send XON or <QUIT>.
E44
The RS-interface parameters are no longer identical for both in-
struments. Re-set.
E45
The receive buffer of the Sample Changer contains an incomplete
command (LF missing). Transmission is therefore blocked.
Exit: send LF or <QUIT>.
E50…E59
I/O-Test error
E60…E82
RS232-Test error
Instrument-specific Errors:
E201
Sample changer function error
E202
Dosimat function error