2.5 Integration
730 Sample Changer, Installation
18
Remote control commands:
CTL:Rm : START device1
starts Titrino 1
CTL:Rm : *************1
"
CTL:Rm : START device2
starts Titrino 2
CTL:Rm : ********1*****
"
CTL:Rm : START device*
starts both Titrinos at the same time
CTL:Rm : ********1****1
"
Scan commands:
SCN:Rm : end1
awaits end of titration of Titrino 1
SCN:Rm : ****1***
"
SCN:Rm : end2
awaits end of titration of Titrino 2
SCN:Rm : *1******
"
SCN:Rm : ready1
awaits readiness of Titrino 1
SCN:Rm : *******1
"
SCN:Rm : ready2
awaits readiness of Titrino 2
SCN:Rm : **1*****
"
SCN:Rm : ready*
awaits readiness of both Titrinos
SCN:Rm : **1****1
"
Operating Titrinos and pH meters at the same time is also possi-
ble in the same manner.
730
B
A
Titrino
C
D
692
712
713
device 1
device 2
Cable 6.2141.030