4.3 Menu Organization
730 Sample Changer, Detailed Description
87
PARAM
4.3.2 Parameters
All settings of the Parameter Menu constitute a method and may be saved as such.
Main Menu:
parameters
number of samples
rack
Number of samples to be processed
1…999,
rack,
∗
rack
= one entire rotation of the rack
*
= infinite number of samples
All sample positions of an engaged rack will be processed when the instrument is on
the 'rack' setting (max. number of rack positions – number of special beakers defined),
while only those positions are counted that contain a sample beaker. It is important
that the changer can recognize the rack. This is only possible when the rack is at the
ground position. It is recommended to initialize the changer with the <CLEAR> key or
<ENDSEQ> and <ENTER> after every rack change.
parameters
>start sequence
Open the submenu with <ENTER>
parameters
>sample sequence
Use <
é
> or <ê> to move up or down one menu
item
parameters
>final sequence
Use <HOME> or <END> to move to first or last
menu item respectively
parameters
>changer settings
Return to the normal state with <QUIT>
parameters
>stirring rates
parameters
>dosing unit def.
parameters
>manual stop
Submenus:
In each of the submenus '>start sequence', '>sample sequence' and '>final sequence'
up to 99 command lines can be entered as a processing sequence. The commands
can be entered directly via the keyboard. The command keys which are situated on the
right half of the keyboard are available.