Metrohm 730 Sample Changer Manuel d'utilisation
Page 37

3.1 Tutorial
730 Sample Changer, Introduction
34
<ENTER>
<ENTER>
•
Now optimize the rinse time in this way. Accept the total time
with <ENTER> and exit the LEARN mode in this way.
Display
9 WAIT 5 s
•
In this line a waiting time is defined that is used here as drip
time.
•
The LEARN mode can also be used with the WAIT command.
<QUIT>
<ê>
<ENTER>
Display:
10 NOP
•
An empty line with a 'NOP' entry (no operation) always forms
the end of a sequence.
•
Exit the sample series with <QUIT> and go to the final se-
quence.
START
•
After all sample beakers have been processed, the end se-
quence is executed.
Display:
>end sequence
1 MOVE 1 : spec.1
•
In this line the special beaker 1 (as conditioning beaker) is put in
front of tower 1. Press <START>.
<START>
Display:
2 LIFT: 1 : work mm
•
In this line lift 1 is put into the operating position with the elec-
trode immersed in the conditioning solution.
•
Press <START>.
QUIT
Display:
3 NOP
•
Now you have reached the end of the end sequence and have
completed the entire run of a sample series.
•
By pressing <QUIT> twice the sample changer returns to the
normal state.
PARAM
7
SAMPLE
START
•
Now prepare some sample beakers and fill the special beaker
with a conditioning solution or water. Place all titrating vessels
on the rack and prepare the Titrino for titration.
•
Enter the number of samples to be processed (<PARAM>) and
define the position of the first sample (SAMPLE =1).
•
Now you can start your first sample series with <START>.