Metrohm 730 Sample Changer Manuel d'utilisation
Page 157

4.10 Remote Control Commands
730 Sample Changer, Detailed Description
154
&Diagnosis …
4.10.2.61
D
iagnose.Init
$G
D
iagnose.Init.Select
param, config,
setup, assembly, all
RAM initialization. Sets all default values for the selected sub-
branch. See also 4.10.2.44.
&Diagnose.Init;$G triggers the initialization.
4.10.2.62
D
iagnose.RamTest
$G
D
iagnose.LcdTest
$G, $S, $H
D
iagnose.ContrastTest
$G, $S
D
iagnose.KeyTest
$G, $S
D
iagnose.IoTest
$G, $S
D
iagnose.RsTest
$G, $S
D
iagnose.EbusTest
$G, $S
D
iagnose.BeeperTest
$G, $S
D
iagnose.RackcodeTest
$G, $S
D
iagnose.FunctionTest
Diagnostic functions. These functions can be triggered with $G.
They may be stopped with $S.
.RamTest
Working memory test
.LcdTest
Display test
.ContrastTest Display contrast test
.KeyTest
Displays key code and function of each key
.IoTest
Remote interface test
.RsTest
(This function is not applicable via RS232 inter-
face)
.EbusTest
External Bus interface test
.BeeperTest
Test of the acoustic signal (only with
Config.Beeper"on", see 4.10.2.14)
.RackcodeTest Test of rack code recognition
.FunctionTest Metrohm internal test
4.10.2.63
D
iagnose.SimulateKey
0
, 1…6, 8…31
Key simulation (Key codes see 4.10.2.37)
4.10.2.64
D
iagnose.InstrNo
Instrument number. This entry is not accessible via RS232 inter-
face.
4.10.2.65
D
iagnose.PowerOn
$G
Simulation of 'Power on'.