5.1 Error Messages

730 Sample Changer, Appendix

162

* dos.unit ## overload

The dosing unit indicated cannot execute a dosing
command. Check burette and piston.

* dos.unit## not ready

The dosing unit selected cannot execute the desired
command because it is busy carrying out another ac-
tion or the current instrument state does not allow it.

* invalid position

The sample position selected is defined as a special
beaker or the special beaker chosen is not defined.

* invalid rack code

The rack code read by the changer could not be
found in the internal position tables.

* missing beaker

After a MOVE command no beaker could be found at
the selected position.

* rack data missing

No sample rack is in position or no rack data can be
found for the sample rack that is in place.

* raise lift first

Turning of a rack could not be carried out because a
lift was below the defined shift position.

* RS232 error ##

The transmission parameters of the RS232 interface
do not agree with those of the receiving instrument.

* user memory full

The memory for the user-defined methods is full.
Before saving a new method, methods that are not
used or used only rarely must be deleted.

* wrong rack

The rack positioned is not the one that was assigned
to the method under 'parameters'.

trap error XXX

Unexpected program error, turn instrument off and on
again.

No display, LEDs tower 1

LCD error (system error 7). Please contact Service.

and tower 2 are lit up

Illegible display when

Wrong 'external bus' address. Set address to 0 and

switching on, LED tower 2 lit

execute RAM-initialization 'assembly' (see page 175).