Metrohm 730 Sample Changer Manuel d'utilisation
Page 145

4.10 Remote Control Commands
730 Sample Changer, Detailed Description
142
&Config …
4.10.2.14
C
onfig.Aux.Language
english
, deutsch, français, español
C
onfig.Aux.Contrast
0…3…7
C
onfig.Aux.Beeper
ON
, OFF
C
onfig.Aux.DevName
8 ASCII characters
C
onfig.Aux.Prog
read only
C
onfig.Aux.MaxLift
0...235...325 mm
C
onfig.Aux.Pumps1
0, 1, 2
C
onfig.Aux.Pumps2
0, 1, 2
C
onfig.Aux.SwingH
on, off
C
onfig.Aux.MonBeak
on
, off
Basic configuration / miscellaneous
Language:
Selection of dialog language
Contrast:
Display contrast in steps from 0 to 15
Beeper:
Acoustic warning signal on/off
DevName:
Instrument name as identification for connections
with other instruments (8 ASCII characters, no spe-
cial characters)
Prog:
Program version ($Q sends "730.0013")
MaxLift:
Max. Lift height i.e. lowest possible lift position
∗
Pumps1:
Numbers of pumps on tower 1
∗
Pumps2:
Numbers of pumps on tower 2
∗
SwingH:
Swing Head on/off
∗
MonBeak:
Beaker Sensor on/off
∗
∗
Modifications effect after RESET or power on.
4.10.2.15
C
onfig.RackDef.RackNo
1
...16
Rack number
4.10.2.16
C
onfig.RackDef.Code
000001
b...111111b
C
onfig.RackDef.Type
8 characters
C
onfig.RackDef.WorkH
0
...325 mm
C
onfig.RackDef.RinseH
0
...325 mm
C
onfig.RackDef.ShiftH
0
...325 mm
C
onfig.RackDef.SpecialH
0
...325 mm
Rack definitions. Depending on the current rack number (see
4.10.2.15) the corresponding one of the 16 possible data sets is
overlayed.
Code:
Identification code of the rack, has to be unique.
Type:
Rack type. Valid entries are the names of the posi-
tion tables, see 4.10.2.19.
WorkH:
Working position of the lift in mm from the upper
stop of the tower.
RinseH:
Rinsing position of the lift in mm from the upper stop
of the tower.
ShiftH:
Shifting position of the lift in mm from the upper
stop of the tower.
SpecialH:
Special position of the lift in mm from the upper stop
of the tower.