4.10 Remote Control Commands
730 Sample Changer, Detailed Description
143
4.10.2.17
C
onfig.RackDef.SpezBeak.1.Pos
0
...number of rack positions
etc. until .8
Rack positions of special beakers 1 to 8 (position 0 = not defined).
4.10.2.18
C
onfig.PosTab.TabIdx
0
...31
Index of position tables.
4.10.2.19
C
onfig.PosTab.Name
8 ASCII characters
C
onfig.PosTab.R1Num
2...(R2Num – 2)
C
onfig.PosTab.R2Num
(R1Num + 2)...(R3Num – 2)
C
onfig.PosTab.R3Num
(R2Num + 2)...200
C
onfig.PosTab.R1Off
0
...3599
C
onfig.PosTab.R2Off
0
...3599
C
onfig.PosTab.Num
1
...200
Definitions of position tables. Depending on the table index (see
4.10.2.18) the corresponding one of the 16 possible data sets is
overlayed.
Name:
Identification of the rack type. Will be available as a
selector under &Config.RackDef.Type (see
4.10.2.16).
R1Num:
Highest Beaker position in row 1
R2Num:
Highest Beaker position in row 2
R3Num:
Highest Beaker position in row 3
R1Off
Offset in
1
/
10
-angular degrees for the beaker posi-
tions in row 1 (for the beaker test)
R2Off
Offset in
1
/
10
-angular degrees for the beaker posi-
tions in row 2 (for the beaker test)
Num:
Number of rack positions
4.10.2.20
C
onfig.PosTab.1.Value
0
...3599
up to .200.Value
Angular offset for the particular rack position in tenth of angular
degrees (
1
/
10
-degrees).
Definition of new rack types:
The definition of rack types is only possible via RS232 interface.
•
Set table index (see 4.10.2.18)
•
Enter name of rack type (&Config.PosTab.Name, see
4.10.2.19)
•
Enter highest beaker position in row 1 to 3
(&Config.PosTab.R1–3Num, see 4.10.2.19)
•
Enter the offset angle between beaker position 1 in row 1 and
beaker position (R1num+1) in row 2, respectively and tower 1
(&Config.PosTab.R1–2Off, see 4.10.2.19)
•
Define number of rack positions (&Config.Pos.Tab.Num, see
4.10.2.19)
•
Enter the particular angular offset for each rack position
(Irregular layouts of rack positions are possible)