4.3 Menu Organization
730 Sample Changer, Detailed Description
85
Position of special beaker 2
>>special positions
special beaker 2
0
0…number of positions
etc. up to special beaker 8
Up to 8 special beaker positions can be defined. More information
about racks and special beakers is found in Chapter 4.5 "Sample
Racks", page 104.
configuration
>dosing units
Submenu for dosing unit settings
Open the submenu with <ENTER>
>dosing units
dosing unit
1
Select the dosing unit
(700 Dosino or 685 Dosimat)
1…12
The number of the dosing (see page 23) unit must be confirmed with
<ENTER>. This will then be displayed in the first menu line.
>dosing units
1
max. rate 1
160 ml/min
Max. dosing speed (dependent
on burette size)
0.01…160 ml/min
>dosing units
1
tube length 1
1000 mm
Length of the tubing on Dosino
port 1
0…1000…30000 mm
>dosing units
1
tube diameter 1 2 mm
Inner diameter of the tubing on
Dosino port 1
0.1…2.0…20 mm
>dosing units
1
max. rate 2
160 ml/min
Max. dosing speed (dependent
on burette size)
0.01…160 ml/min
>dosing units
1
..........until Port 4
Enter the tubing parameters for
all four ports of a Dosino
Only the dosing speed is relevant for the 685 Dosimat, the other
parameters are ignored.
use
<QUIT>
to access the
next highest