4.1 The Display
730 Sample Changer, Detailed Description
59
4 Detailed Description
4.1 The Display
The display consists of two lines, each having 24 characters.
The first line serves as a title line in which the current method and
the sample counter reading are displayed. In edit mode the menu
title is shown.
The second line serves as a status line which displays specific ac-
tivities depending on the operating state. In edit mode it serves as
an entry line.
Normal State
Sample counter к
к
Method name и
и
Pump status и
и
******** counter 1/12
PUMP--++ STIR+--- ready
з
з
Changer status
й
й
Stirrer status
Method Processing
Running sequence и
и
******** counter 2/12
START 03 WAIT 11 s
з
з
Parameter
й
й
Current command with line number
Edit Mode
ê
Menu title
Menu line or command è
>Sample sequence
1 MOVE 1 : Sample
ç
Second parameter
é
First parameter