4.10 Remote Control Commands
730 Sample Changer, Detailed Description
144
4.10.2.21
C
onfig.WetPart.WetPartNo
1
...12
Identification of dosing unit.
4.10.2.22
C
onfig.WetPart.1.MaxRate
0.01...160 ml/min
C
onfig.WetPart.1.Length
0...1000...30000 mm
C
onfig.WetPart.1.Diameter
0.1...2...20 mm
until .4.Diameter
(Default value .2.Length : 250 mm)
Configuration of the dosing units. Depending on the selection of
the dosing unit (see WetPartNo 4.10.2.21), the corresponding data
set is overlayed. These settings are only relevant for Dosinos. For
each port of the Dosino individual settings are possible.
MaxRate:
Max. possible dosing and filling speed allowed
Length:
Tubing length on selected port
Diameter:
Inner tubing diameter on selected port
4.10.2.23
C
onfig.RSset
$G
$G effects all the RS settings. Modifications are only possible in
inactive instrument state. After setting the interface parameters
wait for 2 sec. to allow the components to equilibrate.
4.10.2.24
C
onfig.RSset.Baud
300, 600, 1200, 2400, 4800, 9600
C
onfig.RSset.DataBit
7, 8
C
onfig.RSset.StopBit
1
, 2
C
onfig.RSset.Parity
even, odd, none
C
onfig.RSset.Handsh
HWs
, HWfull, SWchar, SWline, none
Settings for data transmission via RS interface, baud rate, data bit,
stop bit, parity and type of handshake, see also page 155ff.
4.10.2.25
C
onfig.RSset.CharSet
IBM
, HP, Epson, Seiko, Citizen
Setting the character set and print mode. For data communication
with computers select 'IBM' (IBM code page 437).