4.2 The Keyboard
730 Sample Changer, Detailed Description
70
Key
Normal State
Editing
DOS
6
Dosing control
•
The <DOS> key serves for
the control of Dosimats
and Dosinos. These are
controlled via the "External
Bus" connector.
•
The first parameter is for
the selection of the dosing
unit. The second parame-
ter represents the function.
•
In addition to dosing vol-
ume and the fill command,
the functions specific to
the Dosino 'release', 'pre-
pare', 'empty', 'eject', 'ad-
just' and 'level' can also be
executed.
•
The dosing and filling
speed can be set in the
parameter menu or with
the <DEF> key.
Numerical entry ('6')
or
Dosing control
•
In a sequence you can not
only add a certain volume
under process control, it is
also possible to initiate the
burette filling with Dosimats
or initiate the designated
functions with Dosinos as
there are: filling, emptying,
preparation, release,
ejecting, adjusting or level
of the exchange unit.
1
SCAN
Display input signals
•
Display of the incoming
signal or data on the re-
mote or RS interface. This
function serves to control
the data communication or
states of connected in-
struments.
•
The first parameter repre-
sents the interface selec-
tion. The signals for data
being received at that
moment are displayed as
the second parameter.
•
If the parallel remote in-
terface (Rm) is selected,
the signal states of the in-
coming remote lines are
displayed in binary form
(1=line active, 0=line inac-
tive). Further details relat-
ing to this see page 119ff.
Numerical entry ('1')
or
Scan input signals
•
In a sequence the SCAN
command causes method
processing to stop until the
predefined bit pattern (with
the remote interface) or the
given character string (with
the RS232 interface) is re-
ceived.
•
Predefined bit patterns are
available for the remote
interface and can be se-
lected via simple short-
names (e.g. "ready1" or
"end2").