Metrohm 730 Sample Changer Manuel d'utilisation
Page 87

4.3 Menu Organization
730 Sample Changer, Detailed Description
84
>rack definitions
1
code
000001
Identification code of the rack
See table on page 104.
6 bits
The rack code must be unique and can only occur once in the in-
strument.
>rack definitions
1
type:
M12-0
Type description of the rack
See table on page 104.
M12-0…
<SELECT> enables the choice of Metrohm-specific and self-defined
rack types.
>rack definitions
1
work position
0 mm
Working position of the lift (in mm
from the upper stop)
0…325 mm
Pressing <CLEAR> directly accepts the current lift position of the
active tower.
>rack definitions
1
rinse position
0 mm
Rinsing position of the lift (in mm
from the upper stop)
0…325 mm
Pressing <CLEAR> directly accepts the current lift position of the
active tower.
>rack definitions
1
shift position
0 mm
Shifting position of the lifts (in mm
from the upper stop)
0…325 mm
Pressing <CLEAR> directly accepts the current lift position of the
active tower.
>rack definitions
1
special position
0 mm
Position of the lift (in mm from the
upper stop)
0…325 mm
Pressing <CLEAR> directly accepts the current lift position of the
active tower.
>rack definitions
1
>>special positions
Special positions submenu
Open with <ENTER>
Position of special beaker 1
>>special positions
special beaker 1
0
0…number of positions
use
<QUIT>
to access the
next highest
level
use
<QUIT>
to access the
next highest
level
use
<QUIT>
to access the
next highest
level