Metrohm 730 Sample Changer Manuel d'utilisation

Page 87

Advertising
background image

4.3 Menu Organization

730 Sample Changer, Detailed Description

84

>rack definitions

1

code

000001

Identification code of the rack
See table on page 104.

6 bits

The rack code must be unique and can only occur once in the in-
strument.

>rack definitions

1

type:

M12-0

Type description of the rack
See table on page 104.

M12-0

<SELECT> enables the choice of Metrohm-specific and self-defined
rack types.

>rack definitions

1

work position

0 mm

Working position of the lift (in mm
from the upper stop)

0…325 mm

Pressing <CLEAR> directly accepts the current lift position of the
active tower.

>rack definitions

1

rinse position

0 mm

Rinsing position of the lift (in mm
from the upper stop)

0…325 mm

Pressing <CLEAR> directly accepts the current lift position of the
active tower.

>rack definitions

1

shift position

0 mm

Shifting position of the lifts (in mm
from the upper stop)

0…325 mm

Pressing <CLEAR> directly accepts the current lift position of the
active tower.

>rack definitions

1

special position

0 mm

Position of the lift (in mm from the
upper stop)

0…325 mm

Pressing <CLEAR> directly accepts the current lift position of the
active tower.

>rack definitions

1

>>special positions

Special positions submenu
Open with <ENTER>

Position of special beaker 1

>>special positions

special beaker 1

0

0…number of positions

use

<QUIT>

to access the

next highest

level

use

<QUIT>

to access the

next highest

level

use

<QUIT>

to access the

next highest

level