4.10 Remote Control Commands
730 Sample Changer, Detailed Description
139
4.10.2 Description of the Remote Control Commands
&Mode …
4.10.2.1
M
ode
$G, $S, $H, $C
Start ($G) or stop ($S) the current method. Interrupt with $H
(hold), resume with $C (continue).
4.10.2.2
M
ode.Method
read only
Name of current method in the working memory. $Q sends
8 ASCII characters.
∗∗∗∗∗∗∗∗
stands for a blank method.
4.10.2.3
M
ode.SmplNo
1...999,
∗
, rack
Number of samples. This entry determines the number of runs for
the sample sequence.
∗
= infinite number of samples. Processing has to be stopped
with &M;$S or <STOP>.
rack = (Number of rack positions – number of special beakers
defined).
4.10.2.4
M
ode.StartSeq.1.Cmd
NOP
, MOVE, LIFT, STIR,
etc. up to .99
DEF, PUMP, DOS, SCAN,
SAMPLE, CTRL, WAIT, ENDSEQ
Defines the command of the indexed command line in a start se-
quence. The introduction of a command appends the according
sub-branch from &Assembly (see page 136) to the index node.
A NOP-entry cuts the appended sub-branch from the index node.
Each entry at the end of a sequence appends a new node
&Mode.StartSeq.
∗
.Cmd("NOP").
4.10.2.5
M
ode.StartSeq.1.
∗
.Move…, .Lift…, .Stir…, .Pump…,
etc. up to .99
.Dos…, .Scan…, .Ctrl…, .Def…,
.Sample…, .Wait…, .End
Indexed start sequence; its commands will be executed line by line
in processing. At the index node &Mode.StartSeq.1.Cmd (see
4.10.2.4) the according sub-branch from &Assembly… will be ap-
pended depending on the selected command (see 4.10.2.4).
Example:
&Mode.StartSeq.1.Cmd("MOVE")
⇒
&Mode.StartSeq.1.Move.Target("1")
⇒
&Mode.StartSeq.1.Move.Position("sample")