4.6 Dosimats and Dosinos
730 Sample Changer, Detailed Description
113
Release exchange unit
Release exchange unit
DOS: XX : release ml
Prepare Dosino for removing the ex-
change unit.
Before removing the exchange unit the burette must be filled and
the Dosino stop-cock run to the exchange position. Under stan-
dard conditions the volume required for filling the burette is aspi-
rated from the filling port. To fill the burette for example, with dis-
tilled water for storage, this can be changed with the following
commands.
>dosing unit def. XX
rinsing Port Y
or
<DEF> DRIVE.PORT XX.Y : rinse
Ejecting
Ejecting
DOS: XX : eject ml
Empty the Dosino burette.
The contents of the burette is ejected completely via the dosing
port. This can be redefined as desired (see Dosing).
Adjust
Adjust
DOS: XX : adjust ml
Conmpensate the play
The mechanical play between the dosing piston and the spindle is
compensated. This command is important for exact pipetting when
small volumes are aspired into the pipetting tube and ejected
again. First, the content of the cylinder is ejected completely
(eject), then before the solution is aspired, the piston is adjusted
(adjust).
Level
Level
DOS: XX : level ml
Compensate the play
The mechanical play between the dosing piston and the spindle is
compensated after the cock has been turned into the dosing posi-
tion. The dosing port can be redefined as desired (see Dosing).
This command is executed to increase the precision.