4.2 The Keyboard
730 Sample Changer, Detailed Description
65
Key
Normal State
Editing
•
The commands that con-
trol the titration heads and
pumps are always exe-
cuted at the currently ac-
tive tower when operating
in manual mode.
•
With every repeated key-
stroke the next value that
can be selected is dis-
played. The last value is
followed again by the first.
•
The data is accepted with
<ENTER>.
RESET
CLEAR
Initialization of the
changer and the dosing
units
•
The <RESET> key serves
to initialize the changer
and the dosing units.
•
A method in the working
memory remains un-
changed. The sample rack
and the lifts return to their
initial positions in this
case. A 'release' command
will be executed when
Dosinos are connected.
Interruption of a
method after the cur-
rent sequence
•
During processing of a
method, the sample series
can be aborted with
<CLEAR> so that the
sample currently being
processed is processed to
completion. The end se-
quence is not executed in
this case.
Deleting parameters,
setting the default val-
ues
•
The <CLEAR> key sets the
initial (default) value given
for a parameter.
Delete last character
•
In text edit mode the last
character will be deleted
with <CLEAR> (Back-
space).