Metrohm 730 Sample Changer Manuel d'utilisation
Page 114

4.6 Dosimats and Dosinos
730 Sample Changer, Detailed Description
111
The maximum dosing and filling rates that can be entered in the
configuration menu under '>dosing units' for every port of a dosing
unit depend on the burette size:
Volume of the
dosing unit
Max. dosing rate
Resolution
2 mL
7 mL/min
0.2
µ
L
5 mL
17 mL/min
0.5
µ
L
10 mL
33 mL/min
1.0
µ
L
20 mL
67 mL/min
2.0
µ
L
50 mL
160 mL/min
5.0
µ
L
The following commands can be executed with Dosinos. The as-
sociated inlets and outlets (ports) can be defined in the Parameter
Menu under '>dosing unit def.' as standard arrangement for a par-
ticular method or with a DEF command in a process sequence or
manual operation.
Dosing
Dosing
DOS: XX : yyy.yy ml
Dosing a certain volume.
The volume indicated is ejected at the dosing port. The burette is
not refilled after every operation. The dosing port can be redefined
as desired:
>dosing unit def. XX
dosing port Y
or
<DEF>
DRIVE.PORT XX.Y : dos.
Filling
Filling
DOS: XX : fill ml
Filling the Dosino burette.
The burette is completely filled. It is aspirated via the filling port.
This can be redefined as desired:
>dosing unit def. XX
filling port Y
or
<DEF>
DRIVE.PORT XX.Y : fill