4.2 The Keyboard
730 Sample Changer, Detailed Description
62
Key
Normal State
Editing
HOME
Bring lift to zero-
position
•
The <HOME> key runs the
lift of the active tower to
the zero-position (0 mm),
i.e. to the upper stop.
Select the first line of a
menu
•
With the <HOME> key, the
first line in a menu or a se-
quence can be accessed.
•
Any data that has been
altered in a menu or com-
mand line is not carried
over. See <ENTER> key.
END
Lift in work position
•
The <END> key runs the
lift of the active tower into
the work position.
•
The work position is de-
fined separately for every
sample rack in the configu-
ration menu under '>rack
definitions' (in mm from
the rest-position, i.e. as
measured from the upper
stop).
Select the last line of a
menu
•
With the <END> key, the
last line in a menu or a se-
quence can be accessed.
•
Any data that has been
altered in a menu or com-
mand line is not carried
over. See <ENTER> key.
Run lift upwards
•
Run lift of the active tower
of the sample changer
upwards using the <é>
key. The lift movement is
executed as long as the
key remains pressed.
•
The speed of the lift
movement can be ad-
justed separately for each
tower in the Parameter
Menu or with the <DEF>
key.
Select previous menu
line
•
In a Select Menu or a se-
quence the <é> key ac-
cesses the previous line.
•
Any data which has been
altered in a menu or com-
mand line is not carried
over. See <ENTER> key.