Metrohm 730 Sample Changer Manuel d'utilisation
Page 93

4.3 Menu Organization
730 Sample Changer, Detailed Description
90
>changer settings
on beaker error:
MOVE
Defining the reaction to a missing
beaker
MOVE, display
MOVE =
The last action will be exe-
cuted once more. The next
position according to the cur-
rent SAMPLE command will
be chosen.
display =
Processing will be inter-
rupted and a warning dis-
played.
If processing should not be halted when a sample beaker is missing,
'MOVE' can be chosen. In case of a missing beaker another MOVE
command is executed with the next sample. The next sample is cho-
sen according to the last SAMPLE command, i.e. if the previous
command was 'SAMPLE +2', the sample in the rack position after
the next beaker is chosen, etc.
If a special beaker is missing an error message is always displayed
and processing is halted.
parameters
>stirring rates
Submenu for Stirrer Settings
Open the submenu with <ENTER>
>stirring rates
stirrer 1
3
Stirring speed stirrer 1
in intervals from 1 to 15
1…3…15
>stirring rates
stirrer 2
3
Stirring speed stirrer 2
in intervals from 1 to 15
1…3…15
>stirring rates
stirrer 3
3
Stirring speed stirrer 3
in intervals from 1 to 15
1…3…15
>stirring rates
stirrer 4
3
Stirring speed stirrer 4
in intervals from 1 to 15
1…3…15