4.10 Remote Control Commands
730 Sample Changer, Detailed Description
148
4.10.2.40
S
etup.Lock.Keyboard
on, off
S
etup.Lock.Config
on, off
S
etup.Lock.Parameter
on, off
S
etup.Lock.UserMeth.Recall
on, off
S
etup.Lock.UserMeth.Store
on, off
S
etup.Lock.UserMeth.Delete
on, off
S
etup.Lock.Display
on, off
"on" means disable the corresponding function.
.Keyboard
Disables all keys of the keyboard, except the
<START>, STOP> and <CLEAR> key.
.Config
Locks the configuration menu
.Parameter
Locks the parameter menu
.Usermeth.Recall Locks the function "recall method"
.UserMeth.Store Locks the function "store method"
.Usermeth.Delete Locks the function "delete method"
.Display
Disables the LCD display. The Sample
Changer will not support the display.
4.10.2.41
S
etup.Mode.StartWait
on, off
Indefinite start delay. Only for remote control.
4.10.2.42
S
etup.AutoInfo.Status
on
, off
S
etup.AutoInfo.P
on, off
S
etup.AutoInfo.Ch.G
on, off
S
etup.AutoInfo.Ch.GC
on, off
S
etup.AutoInfo.Ch.R
on, off
S
etup.AutoInfo.Ch.S
on, off
S
etup.AutoInfo.Ch.H
on, off
S
etup.AutoInfo.Ch.C
on, off
S
etup.AutoInfo.Ch.B
on, off
S
etup.AutoInfo.Ch.F
on, off
S
etup.AutoInfo.Ch.OM
on, off
S
etup.AutoInfo.Ch.CM
on, off
S
etup.AutoInfo.E
on, off
"on" means, the Sample changer automatically sends a corre-
sponding message when the specified event occurs.
.Status
Enables/disables the preset AutoInfo messages
.P
PowerOn: Simulation of PowerOn (see 4.10.2.43).
Not from mains.
Messages from changer functions:
.Ch.G
Go: Method has been started
.Ch.GC
Go Command: Start command has been received
.Ch.R
Ready: Status 'Ready' reached
.Ch.S
Stop: Status 'Stop' reached
.Ch.H
Hold: Status 'Hold' reached
.Ch.C
Continue: Resuming after Hold
.Ch.B
Begin: Begin of sample sequence
.Ch.F
Final: End of sample sequence