4.10 Remote Control Commands
730 Sample Changer, Detailed Description
140
4.10.2.6
M
ode.SampleSeq.1.Cmd
NOP, MOVE, LIFT, STIR,
etc. up to .99
DEF, PUMP, DOS, SCAN,
SAMPLE, CTRL, WAIT, ENDSEQ
Defines the command of the indexed command line in a sample
sequence. See start sequence (4.10.2.4).
4.10.2.7
M
ode.SampleSeq.1.
∗
.Move…, .Lift…, .Stir…, .Pump…,
etc. up to .99
.Dos…, .Scan…, .Ctrl…, .Def…,
.Sample…, .Wait…, .End
Indexed sample sequence; its commands will be executed line by
line in processing. See start sequence (4.10.2.5).
4.10.2.8
M
ode.FinalSeq.1.Cmd
NOP, MOVE, LIFT, STIR,
etc. up to .99
DEF, PUMP, DOS, SCAN,
SAMPLE, CTRL, WAIT, ENDSEQ
Defines the command of the indexed command line in a final se-
quence. See start sequence (4.10.2.4).
4.10.2.9
M
ode.FinalSeq.1.
∗
.Move…, .Lift…, .Stir…, .Pump…,
etc. up to .99
.Dos…, .Scan…, .Ctrl…, .Def…,
.Sample…, .Wait…, .End
Indexed final sequence; its commands will be executed line by line
in processing. See start sequence (4.10.2.5).
4.10.2.10
M
ode.Changer.RackNo
0
…16
M
ode.Changer.L1Rate
3...25 mm/s
M
ode.Changer.L2Rate
3...25 mm/s
M
ode.Changer.ShRate
3...20 w/s
M
ode.Changer.ShDir
+, –, auto.
M
ode.Changer.BeakTest
single
, both
M
ode.Changer.ModeSample
MOVE
, display
Changer settings.
RackNo:
Rack number, forces the use of the defined rack
with the current method (0 = any rack).
L1Rate:
Lift speed for tower 1, in mm/sec
L2Rate:
Lift speed for tower 2, in mm/sec
ShRate:
Turning speed of the rack, in angular degrees/sec
ShDir:
Turning direction of the rack (ascending or de-
scending rack positions; auto. means automatic
choice of the shortest path)
BeakTest:
Mode for beaker test
(single = test one single (selected) tower, both =
test both towers). The beaker test is executed
right after each MOVE command.
ModeSample: Reaction on missing sample beaker. (MOVE =
next sample beaker will be chosen regarding the
recent SAMPLE command, display = display
warning.)