Metrohm 730 Sample Changer Manuel d'utilisation

Page 86

Advertising
background image

4.3 Menu Organization

730 Sample Changer, Detailed Description

83

The following three entries become effective after a RESET or
switching the changer on again.

>

auxiliaries

max. lift way

235 mm

Max. stroke path for lift 1 and 2

0…235…325 mm

This setting for max. lift way is important for safe operation. If the
value for this entry is correct, breaking the electrode glass can be
avoided because this prevents the titration head from being driven
lower than the position indicated.

The range of 235 mm to 325 mm is only significant for high tower
models. For the normal tower model only values less than 236 mm
may be set.

>auxiliaries

pumps on tower 1

1

Number of pumps on tower 1

0,1,2

>auxiliaries

pumps on tower 2

1

Number of pumps on tower 2

0,1,2

>auxiliaries

swing head

OFF

Switch on/off swing head

ON,OFF

>auxiliaries

beaker sensor

ON

Switch on/off beaker sensor

ON,OFF

configuration

>rack definitions

Submenu for the definition of the individual racks

Open the submenu with <ENTER>

>rack definitions

rack number

1

Number of the rack

1…16

The number of the rack in position is automatically displayed here,
when its configuration is already stored and when a RESET has
been performed. If the configuration of another rack must be
changed, its rack number must be entered and confirmed with
<ENTER>. The rack number will be shown in the first menu line for
the subsequent entries. An overview of the available rack types is
shown on page 104.