Metrohm 730 Sample Changer Manuel d'utilisation
Page 86

4.3 Menu Organization
730 Sample Changer, Detailed Description
83
The following three entries become effective after a RESET or
switching the changer on again.
>
auxiliaries
max. lift way
235 mm
Max. stroke path for lift 1 and 2
0…235…325 mm
This setting for max. lift way is important for safe operation. If the
value for this entry is correct, breaking the electrode glass can be
avoided because this prevents the titration head from being driven
lower than the position indicated.
The range of 235 mm to 325 mm is only significant for high tower
models. For the normal tower model only values less than 236 mm
may be set.
>auxiliaries
pumps on tower 1
1
Number of pumps on tower 1
0,1,2
>auxiliaries
pumps on tower 2
1
Number of pumps on tower 2
0,1,2
>auxiliaries
swing head
OFF
Switch on/off swing head
ON,OFF
>auxiliaries
beaker sensor
ON
Switch on/off beaker sensor
ON,OFF
configuration
>rack definitions
Submenu for the definition of the individual racks
Open the submenu with <ENTER>
>rack definitions
rack number
1
Number of the rack
1…16
The number of the rack in position is automatically displayed here,
when its configuration is already stored and when a RESET has
been performed. If the configuration of another rack must be
changed, its rack number must be entered and confirmed with
<ENTER>. The rack number will be shown in the first menu line for
the subsequent entries. An overview of the available rack types is
shown on page 104.
use
<QUIT>
to access the
next highest
level