4.8 Remote Interface
730 Sample Changer, Detailed Description
122
CTL Command
Parameter
Bit Pattern
Function
INIT
00000000000000
initializes the remote interface *
)
START device1
*************1
starts device1 (for ex., Titrino, Titroprocessor...)*
)
START device2
********1*****
starts device2 (see above, only with multi-cable)
START device*
********1****1
starts devices1 and 2 " *
)
START dos1
*******1******
starts Dosimat on device1 (Titrino via "activate")
START dos2
*****1********
starts Dosimat on device2 "
START dos*
*****1*1******
starts Dosimat on device1 and 2 "
METER mode pH
*********0001*
switches 692 Ion Meter or 691,713 pH Meter to
pH measurement
METER mode T
*********0010*
switches 692 Ion Meter or 691, 713 pH Meter to
temperature measurement
METER mode U
*********0011*
switches 692 Ion Meter or 691, 713 pH Meter to
mV-measurement
METER mode I
*********0100*
switches 692 Ion Meter or 691, 713 pH Meter to
Ipol (mV-measurement)
METER mode C
*********1000*
switches 692 Ion Meter to conc. measurement
METER cal pH
*********0101*
switches 692 Ion Meter or 691, 713 pH Meter to
pH-calibration
METER cal C
*********1001*
switches 692 Ion Meter to conc. calibration
METER enter
*********1111*
simulates the <ENTER> key on 692 Ion Meter or
691, 713 pH Meter (obligatory with the 691 for
pH calibration in order to start measurement of
the second buffer)
With the START commands a short pulse (200 ms) is emitted.
*
)
with pH meters and Ion meters a result print-out is provoked
Manual stop options
Parameter
Bit Pattern
Function
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
STOP device1
************1*
stops device1 (for ex., Titrino, Titroprocessor...)
STOP device2
******1*******
stops device2 (see above, only with multi-cable)
STOP device*
******1*****1*
stops devices1 and 2 "
With the STOP commands a short pulse (200 ms) is emitted.