4.8 Remote Interface
730 Sample Changer, Detailed Description
123
SCN Command
Parameter
Bit Pattern
Function
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
ready1
*******1
queries "ready" state of device1 (Titrinos)
ready2
**1*****
queries "ready" state of device2 "
ready*
**1****1
queries “ready" state of device1 and 2 "
end1
****1***
expects the end pulse of device1 (for ex. EOD)
end2
*1******
expects the end pulse of device2 “
endmeter
***11***
expects the end pulse from 692 Ion Meter or
713 pH Meter (during the waiting interval, the
stirrer 1 is turned on)
With the 'ready
∗
' parameter the readiness of two concurrently operating instruments
may be queried. The 'ready' line of both instruments must be automatically set (static)
at the end of a determination. Two instruments that emit short pulses at the end of e.g.
measurement, can not be controlled at the same time.