Config – Metrohm 730 Sample Changer Manuel d'utilisation
Page 135

4.10 Remote Control Commands
730 Sample Changer, Detailed Description
132
&Config
Object
Description
Input range
Reference
& Root
:
Ã
C
onfig
Instrument
configuration
-
4.10.2.14
: Ã
.Aux
Miscellaneous
³ Ã
.Language
Dialog language
english
, german
³ ³
français, español
³ Ã
.Contrast
Display contrast
0...3...7
³ Ã
.Beeper
Beeper on/off
on
, off
³ Ã
.DevName
Instrument identification
8 ASCII characters
³ Ã
.Prog
Program version
read only
³ Ã
.MaxLift
Max. Lift height
0...235...320 mm
³ Ã
.Pumps1
Number of pumps lift 1
0, 1, 2
³ Ã
.Pumps2
Number of pumps lift 2
0, 1, 2
³ Ã
.SwingH
Swing Head on/off
on, off
³ À
.MonBeak
Beaker sensor on/off
on
, off
³
Ã
.RackDef
Rack definitions
-
4.10.2.15
³ Ã
.RackNo
Rack number
1
...16
³
∗
³ Ã
.Code
Rack code
000001
...111111(b) 4.10.2.16
³ Ã
.Type
Rack type
M12-0
, M14-0, M16-0,
³ ³
M18-0, M24-0, M48-0,.....
³ Ã
.WorkH
Work position
0
…325 mm
³ Ã
.RinseH
Rinse position
0
…325 mm
³ Ã
.ShiftH
Shift position
0
…325 mm
³ Ã
.SpecialH
Special position
0
…325 mm
³ À
.SpezBeak
Special beaker positions
-
4.10.2.17
³
Ã
.1
Special beaker 1
-
³
: À
.Pos
Beaker position
0
…number of pos.
³
À
.8
Special beaker 8
-
³
À
.Pos
Beaker position
0
…number of pos.
³
Ã
.PosTab
Position table
-
4.10.2.18
³ Ã
.Idx
Index of the table
0
…31
³
∗
³ Ã
.Name
Name of the position table
8 ASCII characters 4.10.2.19
³ Ã
.R1Num
Highest Beaker pos. in row 1
2...(R2Num – 2)
³ Ã
.R2Num
Highest Beaker pos. in row 2
(R1Num + 2)...(R3Num – 2)
³ Ã
.R3Num
Highest Beaker pos. in row 3
(R2Num + 2)...200
³ Ã
.R1Off
Offset in
1
/
10
-ang. degr. for R.1 0...3599
³ Ã
.R2Off
Offset in
1
/
10
-ang. degr. for R.2 0...3599
³ Ã
.Num
Number of positions
1
…200
³ Ã
.1
Position 1
-
4.10.2.20
³ ³ À
.Value
Position in
1
/
10
-angular degr.
0
…3599
³ :
³ À
.12, 14, 16, 24, 48, i.e. .PosTab.Num
:
À
.Value
Position in
1
/
10
-angular degr.
0
…3599