Error list, Error messages – Ohaus I10 Indicator Manuel d'utilisation
Page 65

65
Troubleshooting
9
9
9
9
9
Error Messages
Error Messages
Error Messages
Error Messages
Error Messages
DISPLAYED
MESSAGE
DESCRIPTION
LobAt
Batteries are too weak to display weight values.
UnCAL
Invalid checksum of non volatile memory; unit must be set up for
program constants and calibrated. If error remains, it indicates a
hardware problem. Checksum is verified each time the
I
10 is
turned on.
ErrLo
Underload condition. Dead load weight is less than dead load
during calibration.
ErrHi
Weight is beyond 105% of full scale capacity.
PCErr
Parts counting error.
Err0
Invalid memory checksum; indicates hardware error. Checksum is
verified each time the
I
10 is turned on.
Err2
Indicates overrange of A/D converter. Usually caused by over
weight condition beyond“ErrHi”.
Err3
RS232 Communication error - Invalid command sent.
Err5
Internal value is too large to display. Check Program Constants
menu or Custom Units Conversion Factor.
Err7
tohi
Calibration/Linearization Error. Weight may have been added
during calibration procedure when no load should be on load
receiver.
toLo
Calibration/Linearization Error. Incorrect weight placed on load
receiver during calibration procedure.