Auto-zero tracking – Ohaus I10 Indicator Manuel d'utilisation
Page 24

24
Auto-Zero Tracking
Auto-Zero Tracking
Auto-Zero Tracking
Auto-Zero Tracking
Auto-Zero Tracking
Auto-Zero Tracking minimizes the ef-
fects of temperature changes and shift
of the zero reading. Use this parameter
to select a threshold level (in divisions)
which the
I
10 will consider to be a valid
change in load.
Auto-Zero Tracking occurs only when
the display indicates zero regardless of
the load on the load receiver.
☞
0.6 divisions is normally used for
bench, counter and livestock
scales; 3 divisions for vehicle axle
load and railway scales; and 1.0
division for most other scales.
Auto-Zero may be set to At (Off), 0.6, 1
or 3 divisions as follows:
1.
Access the “AZt” parameter if it is
not already displayed.
2.
Press
until the desired set-
ting is displayed.
To accept the displayed setting and pro-
ceed to the next parameter (Zero),
press
.
To exit the Program Constants menu
and store your settings, repeatedly
press
until SEtUP is displayed,
then press
again to return to the
weighing mode.
At*
Turns feature OFF
0.6
sets threshold to 0.6 divisions
1.0
sets threshold to 1.0 divisions
3.0
sets threshold to 3.0 divisions
* Not available for Legal For Trade
installations.
(OFF)
3
3
3
3
3
Setting Program Constants