Load receiver connections – Ohaus I10 Indicator Manuel d'utilisation
Page 10

10
Load Receiver Connections
Load Receiver Connections
Load Receiver Connections
Load Receiver Connections
Load Receiver Connections
The load receiver may be connected to
the
I
10 by hardwiring the connecting
cable to the Terminal Block.
Terminal Block
Connections
Strain Relief
Hardwiring To The Terminal Block
Hardwiring To The Terminal Block
Hardwiring To The Terminal Block
Hardwiring To The Terminal Block
Hardwiring To The Terminal Block
1.
Pass the cable from the load re-
ceiver through the opening in the
left strain relief.
2.
Connect the wires to the Terminal
Block according to the markings on
the Security Plate Label. If an Ohaus
platform is being used, the leads
are color coded as follows:
Black:
-Excitation
Green:
+Excitation
Red:
+Signal
White:
-Signal
Shield: Ground
Yellow: unused*
Blue:
unused*
* Terminate yellow and blue wires so
conductors do not short out against
shield or each other.