Chapter 5:
Chapter 5:
Chapter 5:
Chapter 5:
Chapter 5:
Calibration
Calibration
Calibration
Calibration
Calibration
The
I
10 must be calibrated after
Program Constants have been ini-
tially set, and when any of the
Program Constants are changed
stants). Calibration is accessed
from the Setup mode.
Before beginning calibration you
will need to have the following
items on hand:
•
a calibration weight equal to the
Calibration Point entered in the
Program Constants menu (see
•
If linearization will be performed,
a weight equal to 1/2 of the
Calibration Point will also be
required
•
the screwdriver which was pro-
vided with the
I
10 (not required
for factory calibrated bench
scales; -C0 models).
☞
Only 2 point calibration is
available for Legal For Trade
installations.
The rear cover of the
I
10 must be
removed before calibrating if the
SETUP switch is set to OFF. Refer
to
f needed.
MODEL
CAL WT. (lbs)
LB30-C0
30
LB60-C0
60
PL100-C0
100
PL150-C0
150
WB10-C0
10
WB25-C0
25
WB50-C0
50
FACTORY CALIBRATED
FACTORY CALIBRATED
FACTORY CALIBRATED
FACTORY CALIBRATED
FACTORY CALIBRATED
BENCH SCALES
BENCH SCALES
BENCH SCALES
BENCH SCALES
BENCH SCALES
If your
I
10 is part of a factory calibrated
bench scale setup (-C0 model number
suffix), the weights listed below are
recommended to perform calibration.
5
5
5
5
5
Calibration