Calibration, Calibration procedure – Ohaus I10 Indicator Manuel d'utilisation
Page 46

46
Calibration Procedure
Calibration Procedure
Calibration Procedure
Calibration Procedure
Calibration Procedure
Make sure there is no load on the plat-
form.
1.
Press and hold
to access the
Setup mode.
Release
when “SEtUP” is
displayed.
2.
Press and hold
until “CAL” is
displayed, then release it.
Dead Load Adjustment
Dead Load Adjustment
Dead Load Adjustment
Dead Load Adjustment
Dead Load Adjustment
3.
When
is released,
“dLOAd” will be displayed, then the
current dead load setting.
4.
If dead load is “good”, then “CAL 0”
Adjustment
Potentiometer
will be displayed. Proceed to step 5.
If dead load is displayed as a per-
centage, for example “dL3.5%”, it
must be adjusted to be between
0.5% and 3.0% as follows:
Using the screwdriver provided, turn
the dead load adjustment potenti-
ometer clockwise to increase
clockwise to increase
clockwise to increase
clockwise to increase
clockwise to increase or
counterclockwise to decrease
counterclockwise to decrease
counterclockwise to decrease
counterclockwise to decrease
counterclockwise to decrease the
displayed value.
Press
to enter the dead load.
☞
If deadload reads “ErrLo”, turn
potentiometer clockwise
clockwise
clockwise
clockwise
clockwise until per-
cent sign (%) appears, then adjust
between 0.5% and 3.0%.
5
5
5
5
5
Calibration