Ohaus I10 Indicator Manuel d'utilisation
Page 47

47
Calibration
5
5
5
5
5
Calibration
Calibration
Calibration
Calibration
Calibration
5.
When “CAL 0” is displayed, press
.
“buSY” will be displayed briefly fol-
lowed by “CAL F” if 2 point calibra-
tion is being performed, or “CAL L”
if 3 point linearization is being per-
formed.
6.
For 2 point calibration, place a weight
equal to the Calibration Point on the
load receiver, then press
and
proceed to step 9.
For 3 point linearization, place a
weight equal to 1/2 the Calibration
Point on the load receiver, then
press
and continue with step
7.
7.
“buSY” will be displayed briefly, then
“CAL F” will be displayed.
8.
Place a weight equal to the Calibra-
tion Point on the load receiver, then
press
.
9.
“buSY” will be displayed briefly, then
“good”, then “SEtUP”.
When “SEtUP” is displayed, the
I
10 is
calibrated.
Press
to return to the weighing
mode.
or