Full Scale Point
Full Scale Point
Full Scale Point
Full Scale Point
Full Scale Point
The maximum capacity of the load re-
ceiver is entered for Full Scale Point
using the same weighing unit selected
for the Calibration Unit.
☞
If Legal for Trade is selected, reso-
lution will be limited to 10,000 divi-
sions. If you attempt to enter a full
scale point that will yield a resolu-
tion greater than 1:10,000, the dis-
play will flash “LFT”, then “Err”
(error), and return to the Gradua-
tions parameter.
To enter the Full Scale Point:
1.
Access the “FS” parameter if it is not
already displayed.
2.
The display will show the current full
scale point with the first digit(s) flash-
ing.
3.
To change the flashing digit(s), re-
peatedly press
until the de-
sired digit is displayed.
To shift to the next digit, press
.
To turn ALL digits ON and reset
them to zero, press
.
4.
Set all digits until the desired Full
Scale Point is displayed.
Divisions = FS Pt/Graduations