Connection to the interface is made us-
ing the 8-pin DIN receptacle on the rear
of the
I
10. The pinout and pin connec-
tions are shown on the security plate
label.
☞
The
I
10 will not output data un-
less pin 3 is held in an ON state
(+3 to +15 VDC). Interfaces not
utilizing the CTS handshake may
tie pin 3 to pin 1 to defeat it.
RS232
Connector
Output Formats
Output Formats
Output Formats
Output Formats
Output Formats
Data output can be initiated in one of
three ways: 1) by pressing PRINT;
2) using the Auto Print feature; 3) send-
ing a “P” command from the computer.
The output format is illustrated in the
RS232 command table (under “P”) which
follows.
RS232 Commands
RS232 Commands
RS232 Commands
RS232 Commands
RS232 Commands
All communication is accomplished us-
ing standard ASCII format. Only the char-
acters shown in the following table are
acknowledged by the
I
10. Any other
commands, control characters or spaces
are ignored.
Commands sent to the
I
10 must be
terminated with a carriage return (CR) or
carriage return-line feed (CRLF). For
example, a tare command should ap-
pear as shown in the adjacent diagram.
Data output by the
I
10 is always termi-
nated with a carriage return-line feed
(CRLF).
TARE COMMAND
TARE COMMAND
TARE COMMAND
TARE COMMAND
TARE COMMAND
Field:
T
CR
LF
Length:
1
1
1
Communication Parameter Settings
Communication Parameter Settings
Communication Parameter Settings
Communication Parameter Settings
Communication Parameter Settings
Baud Rate:
300, 1200 or 9600
selectable in Print menu
Start Bit:
1
Data Bits:
7
Parity:
None
Stop Bits:
2
RS232 Interface