Parts counting – Ohaus I10 Indicator Manuel d'utilisation
Page 51

51
Parts Counting
Parts Counting
Parts Counting
Parts Counting
Parts Counting
Parts Counting mode must first be en-
abled in the Unit Selection menu before
it can be used.
i
n
Chapter 4.
In this mode, the
I
10 calculates and
displays the quantity of parts you place
on the load receiver. Since the indicator
determines the quantity based on the
average piece weight of a single part, all
parts must be reasonably uniform in
weight.
To perform parts counting, use the fol-
lowing procedure:
☞
If you will be using a container to
hold parts on the load receiver,
“PC Err” in the Unit Selection menu
must be set to OFF.
1.
Repeatedly press
until “Con”
is displayed along with the parts
counting indicator.
2.
If a container is being used, place it
on the load receiver.
3.
Press
.
4.
When
is released, “Add05”
will be displayed.
The indicator needs a sample of the
parts to use as a reference for count-
ing. The factory setting for the
sample size is 5 parts, but this can
be changed to any value between 5
and 99.
Weighing, Parts Counting and Printing
6
6
6
6
6