Metrohm 799 GPT Titrino Manuel d'utilisation

Page 178

Advertising
background image

3.2. Remote control commands

799 GPT Titrino

174

L3

EOD (End Of Determination)

L4

Can be set in TIP

L5 Error

L6

Activate pulse + can be set in TIP

L7

Buret volume pulses

L8, 9

---

L10-13 Monitoring in modes STAT, DOS, DOC

.ResetLines: Lines are set to the inactive status (= high).

3.2.2.158. Assembly.Stirrer ON, OFF

Switching stirrer ON/OFF.

3.2.2.159. Setup.Comport

1, 2, 1&2

Selects the Titrino COM for the output of automatic info:

&Setup.Keycode

&Setup.Trace

&Setup.SendMeas

&Setup.AutoInfo

3.2.2.160. Setup.Keycode ON, OFF

ON means the key code of a key pressed on the Titrino is outputted. The key

code comprises 2 ASCII characters; table of the keys with their code, see

page 195. A keystroke of key 11 is sent as follows:

#11

The beginning of the message is marked by a space (ASCII 32).

3.2.2.161. Setup.Tree.Short ON, OFF

Setup.Tree.ChangedOnly ON, OFF

Definition of the type of answer to $Q.

.Short:

With "ON", each path is sent with only the necessary amount of

characters in order to be unequivocal (printed in bold in this manual). A com-

bination of .Short and .ChangedOnly is not possible.

.ChangedOnly:

Sends only the changed values, i.e. values which have

been edited. All paths are sent absolute, i.e. from the root.

3.2.2.162. Setup.Trace ON, OFF

The Titrino automatically reports when a value has been confirmed with <en-

ter> at the Titrino. Message, e.g.:

&SmplData.OFFSilo.Id1"Trace"

The beginning of the message is marked by a space (ASCII 32).

3.2.2.163. Setup.Lock.Keyboard ON, OFF

Setup.Lock.Config ON, OFF

Setup.Lock.Parameter ON, OFF

Setup.Lock.SmplData ON, OFF

Advertising