3.2. Remote control commands

799 GPT Titrino

160

3.2.2.87. MemoryCard.FreeMem read

only

Free memory on the card. $Q sends number of free bytes, e.g. "4928".

3.2.2.88. MemoryCard.BatteryChange $G

MemoryCard.BatteryChange.Date YYYY-MM-DD

Date for changing battery. The date will be set with "$G".

3.2.2.89. MemoryCard.List.Card.1.Name read

only

MemoryCard.List.Card.1.Bytes read

only

for each directory

List of all directories on the memory card with the following information:

.Name:

Name of the directory

.Bytes:

Number of bytes used by the directory

3.2.2.90. MemoryCard.List.ActDir.1.Name read

only

MemoryCard.List.ActDir.1.Mode read

only

MemoryCard.List.ActDir.1.Quantity read

only

MemoryCard.List.ActDir.1.DosUnit read

only

MemoryCard.List.ActDir.1.Bytes read

only

MemoryCard.List.ActDir.1.Checksum read

only

for each method

List of all methods of the current card directory with the following information:

.Name:

Name of the method

.Mode: Mode

.Quantity: Measured

quantity

.DosUnit: Buret of the method

.Bytes:

Number of bytes used by the method

.Checksum: Checksum of the method, see 3.2.2.133.

3.2.2.91. Config.Monitoring.Validation.Status ON,

OFF

Config.Monitoring.Validation.Interval 1...365...9999

Config.Monitoring.Validation.Counter

0...9999

Config.Monitoring.Validation.ClearCount $G

Monitoring of validation.

.Interval: Time interval in days for validation.

.Counter: Time counter in days since last validation.

.ClearCount: Clears the above counter.

3.2.2.92. Config.Monitoring.Calibration.Status ON,

OFF

Config.Monitoring.Calibration.MeasInput

1, 2, diff

Config.Monitoring.Calibration.Interval 1...7...9999

Config.Monitoring.Calibration.Counter

0...9999

Monitoring of pH calibration.

.MeasInput:Measuring input.

.Interval: Time interval in days for calibration from the last calibration date.

.Counter: Time counter in days since last calibration. The counter is reset to

zero if a new calibration is carried out or if the calibration for the

corresponding measuring input is entered manually.