3.2 Remote control commands

799 GPT Titrino

171

pH,U,Ipol mV/s

Upol (uA/s)

T (°C/s)

0.7957

0.7957

-

0.7957

0.7957

-

0.7957

0.7957

-

0.7957

0.7957

-

0.7957

0.7957

-

-

-

-

0.7957

0.7957

0.7957

0.7957

-

-

dMeasdV

(mV/ul)

- - 10.6326 10.6326 10.6326 - - -

ERC

34

- - - - - - -

T

(°C)

25.9 25.9 -

-

25.9 25.9 25.9 -

NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal

drift is NV.

OV will be sent for "overrange".

A time frame can be set up from the cycle number and the cycle time (see

3.2.2.149). The cycle number is set to 0 at the start of a method and it is in-

cremented until the end of the method.

3.2.2.145. Info.ActualInfo.MeasPt.Index read

only

Info.ActualInfo.MeasPt.X read

only

Info.ActualInfo.MeasPt.Y read

only

Info.ActualInfo.MeasPt.Z1 read

only

Info.ActualInfo.MeasPt.Z2 read

only

Info.ActualInfo.EP.Index read

only

Info.ActualInfo.EP.X read

only

Info.ActualInfo.EP.Y read

only

$Q sends the last entry into the measuring point list (.MeasPt) or the last entry

into the list of EP's with DET, MET, SET, KFT.

.MeasPt.X"165" Volume (DET, MET), time (STAT, DOS, DOC, SET, KFT,

MEAS), resp. of the MPList in s

.MeasPt.Y"3.654" Measured value (DET, MET), volume (STAT, DOS, DOC,

SET, KFT), resp. of the MPList in mL

.MeasPt.Z1"6.34" Measured value (STAT, DOS, DOC, SET, KFT, MEAS) of the

MPList, format depends on the measured quantity

.MeasPt.Z2"25.8" Temperature of the MPList in °C

.EP.X"1.234" Volume coordinate of the EP

.EP.Y"5.34" Measured value coordinate of the EP

3.2.2.146. Info.ActualInfo.Oven.HeatTime read

only

Info.ActualInfo.Oven.SampleTemp read

only

Info.ActualInfo.Oven.LowTemp read

only

Info.ActualInfo.Oven.HighTemp read

only

Info.ActualInfo.Oven.GasFlow read

only

Info.ActualInfo.Oven.UnitFlow read

only

$Q sends the current values from a connected KF Oven. If no Oven is con-

nected, the values are empty.

.HeatTime: Heating time of sample in s.

.SampleTemp: Nominal sample temperature in °C.

.LowTemp: Lowest temperature during the sample heating time in °C.

.HighTemp: Highest temperature during the sample heating time in °C.

.GasFlow: Average gas flow during sample heating time.

.UnitFlow: Unit of gas flow.

3.2.2.147. Info.ActualInfo.Display.L1

up to 32 ASCII characters

Info.ActualInfo.Display.L8

up to 32 ASCII characters

Info.ActualInfo.Display.DelAll $G