3.2 Remote control commands
799 GPT Titrino
153
meas, temp, rate, all, none
Mode.Parameter.Monitoring.L10Output.Limit upper,lower,any
Mode.Parameter.Monitoring.L10Output.Output active,
pulse
Parameters for STAT, DOS, DOC: Output of a signal on L10 output (pin 8) of
the remote socket on limit value violation.
The output can be assigned a definite or all monitoring value(s) as well as the
violated limit and the type of signal to be outputted determined.
3.2.2.51. Mode.Parameter.Presel.RateDisp ON,
OFF
Parameters for STAT, DOC: Display of the current rate during the determina-
tion.
3.2.2.52. Mode.Parameter.DosPara.Type volume, time, rate
Mode.Parameter.DosPara.Volume.Volume 0.000...10...99999.99
Mode.Parameter.DosPara.Volume.DisType time, rate
Mode.Parameter.DosPara.Volume.Rate 0.001...150,
max.
Mode.Parameter.DosPara.Volume.Time 1...100...999 999
Mode.Parameter.DosPara.Time.Time 1...100...999 999
Mode.Parameter.DosPara.Time.DisType volume, rate
Mode.Parameter.DosPara.Time.Rate 0.001...150,
max.
Mode.Parameter.DosPara.Time.Volume 0.000...10...99999.99
Mode.Parameter.DosPara.Rate.Rate 0.001...150,
max.
Mode.Parameter.DosPara.Rate.StopType volume, time
Mode.Parameter.DosPara.Rate.Time 1...100...999 999
Mode.Parameter.DosPara.Rate.Volume 0.000...10...99999.99
Parameters for DOS: In the equation Rate = Volume/Time, two parameters
are preset, the third is calculated. Depending on the entry of the type, the fol-
lowing information is required: .Volume, .Time or .Rate.
Entries: .Volume in ml, .Time in s, .Rate in ml/min.
3.2.2.53. Mode.Parameter.DosPara.Pause
0...999 999
Parameters for DOS: Pause time in s.
3.2.2.54. Mode.Parameter.DosPara.TDelta 1...10...999 999
Parameters for DOS, DOC: Time interval for the entry of a measured point in
the list of measured points.
3.2.2.55. Mode.Parameter.DosPara.Temp -170.0...25.0...500.0
Parameters for DOS, DOC: Temperature in °C. The temperature is measured
continuously and updated if a T sensor is attached.
The temperature is used for temperature compensation in pH measurements.
3.2.2.56. Mode.Parameter.Monitoring.MeasVal.MeasModepH,U,Ipol,Upol,OFF
Mode.Parameter.Monitoring.MeasVal.MeasInput
1, 2, diff.
Mode.Parameter.Monitoring.MeasVal.PCurrent 0...1...±127
Mode.Parameter.Monitoring.MeasVal.PVoltage 0...400...±1270
Mode.Parameter.Monitoring.MeasVal.PolElectrTest ON, OFF