3.2 Remote control commands

799 GPT Titrino

149

3.2.2.28. Mode.Parameter.Presel.ActPuls ON, OFF

for SET, KFT: first, all, cond., OFF

Output of a pulse on the I/O line "Activate", see page 224.


3.2.2.29. Mode.Parameter.SET1.EP pH:

0...±20.00, OFF

U,

Ipol:

0...±2000, OFF

Upol:

0...±200.0, OFF

Mode.Parameter.SET1.UnitEp read

only

Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and

Ipol) resp. uA (with Upol). The corresponding unit can be read with .UnitEP.

If the value is on "OFF", no further nodes will appear from SET1.

3.2.2.30. Mode.Parameter.SET1.Dyn pH:

0.01...20.00, OFF

U,

Ipol:

1...2000, OFF

Upol:

0.1...200.0, OFF

Mode.Parameter.SET1.UnitDyn read

only

Mode.Parameter.SET1.MaxRate 0.01...10...150, max.

Mode.Parameter.SET1.MinRate 0.01...25.0...999.9

Parameters for SET: Control parameters, see page 31.

.Dyn:

Dynamics, control range in pH, mV (with U and Ipol) or uA (with

Upol). The corresponding unit can be read with .UnitDyn.

.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum

possible rate with the Exchange Unit in current use.

.MinRate: Minimum titration rate in ul/min.

3.2.2.31. Mode.Parameter.SET1.Stop.Type drift, time

Mode.Parameter.SET1.Stop.Drift 1...20...999

Mode.Parameter.SET1.Stop.Time 0...10...999, inf

Mode.Parameter.SET1.Stop.StopT 0...99

999, OFF

Parameters for SET: Type and size of the stop criterion of the titration.

.Type: Type of stop criterion after stop drift or switch-off delay time.

.Drift: Stop drift in ul/min. Applies when "drift" has been selected.

.Time: Switch-off delay time in s. Applies when "time" has been selected. "inf"

means infinite.

.StopT: Stop time in s. Applies when "time" has been selected and the value of

.Time is set to "inf.".

3.2.2.32. Mode.Parameter.TitrPara.Direction +,

-, auto

Parameters for SET, KFT, STAT: Titration direction.

"auto" means the titration direction is determined automatically by the instru-

ment. If 2 EP's have been set in a SET titration, the titration direction is given

by the two EP's. The entry of the titration direction is then invalid.

3.2.2.33. Mode.Parameter.TitrPara.XPause

0...999 999

Parameter for SET, KFT: Pause time in s. Runs before dosing the start vol-

ume.