3.2 Remote control commands

799 GPT Titrino

145

3.2.2.10. Mode.Def.TempVar.C70

RSX, EPX, CXX

etc. up to .C79

Assignment of temporary variables in a submethod for calculations in TIP.

3.2.2.11. Mode.CFmla

Mode.CFmla.1.Value

0...±999 999

Mode.CFmla.2.Value

etc., up to .19

Calculation constants specific to a method. Stored in the method memory of

the Titrino. Operands specific to the sample (3.2.2.117 and 3.2.2.135) and

values of common variables (3.2.2.109) on the other hand are not stored with

the methods.

3.2.2.12. Mode.Parameter.TitrPara.MptDensity 0...4...9

Mode.Parameter.TitrPara.MinIncr 0...10.0...999.9

Mode.Parameter.TitrPara.VStep 0...0.10...9.999

.MptDensity: Parameter for DET: Measuring point density.

.MinIncr: Parameter for DET: Minimum increment in uL. If the minimum in-

crement is set to 0, measured values are stored vs. time.

.VStep:

Parameter for MET: Volume increment in mL. With "0", there is no

dispensing and measured values vs. time are entered in the

measuring point list.


3.2.2.13. Mode

.

Parameter.TitrPara.DosRate 0.01...150, max.

Parameters for DET and MET: Dispensing rate for the volume increments in

mL/min. Max. means maximum possible dispensing rate with the Exchange

Unit in current use.

3.2.2.14. Mode.Parameter.TitrPara.SignalDrift pH,U,Ipol:0.5...50...999, OFF

Upol:

0.05...50...99.9, OFF

Mode.Parameter.TitrPara.UnitSigDrift read

only

Mode.Parameter.TitrPara.EquTime 0...26...9999, OFF

Parameters for DET and MET: Criteria for the measured value acquisition.

Measured value drift in mV/min (with pH, U, Ipol) or uA/min (with Upol),

equilibration time in s. OFF means that the corresponding criterion is switched

off. If both criteria are OFF, the measured values are acquired immediately af-

ter dispensing.

If the equilibration time has never been edited, it is automatically calculated by

the instrument to match the drift, see page 20. After it has been edited once, it

remains in force with the set value.

3.2.2.15. Mode.Parameter.TitrPara.StartV.Type abs.,

rel.,

OFF

Mode.Parameter.TitrPara.StartV.V 0...999.99

Mode.Parameter.TitrPara.StartV.Factor

0...±999 999

Mode.Parameter.TitrPara.StartV.Rate 0.01...150, max.

Parameters for DET, MET, SET, KFT, STAT: Start volume.