Metrohm 799 GPT Titrino Manuel d'utilisation

Page 156

Advertising
background image

3.2. Remote control commands

799 GPT Titrino

152

3.2.2.46. Mode.Parameter.Evaluation.FixTime.1.Value 0.01...1.00,

OFF

Parameters for STAT: Interpolation of the time at preset fraction of the final

volume. Entry as part of V(tot). The interpolated volumes are available as vari-

able C6X.

3.2.2.47. Mode.Parameter.Monitoring.MeasVal.Status ON,

OFF

Mode.Parameter.Monitoring.MeasVal.LowLim pH: 0...±20.00

Mode.Parameter.Monitoring.MeasVal.UpLim U,

Ipol: 0...±2000

Upol:

0...±200.0

Mode.Parameter.Monitoring.MeasVal.Action end, hold, wait, none

Parameters for STAT, DOC: Monitoring of limit values. Measured values (pH,

U or I) are entered in the list of measured points only if the monitoring is ac-

tive.

On limit value violation, an action can be selected:

end: Abort of the determination.

hold: Interrupt dosing until the sequence is continued by manual switching

(&m $C).

wait: Interrupt dosing until the limit value is no longer violated, then continue

automatically.

3.2.2.48. Mode.Parameter.Monitoring.Rate.Status ON,

OFF

Mode.Parameter.Monitoring.Rate.LowLim 0.000...150

Mode.Parameter.Monitoring.Rate.UpLim 0.000...150

Mode.Parameter.Monitoring.Rate.Action

end, hold, wait, none

Parameters for STAT: Monitoring of limit values of the rate. Entry in ml/min.

On limit value violation, an action can be selected:

end:

Abort of the determination.

hold:

Interrupt dosing until the sequence is continued by manual

switching (& m $C).

wait:

Interrupt dosing until the limit value is no longer violated, then

continue automatically. On violation of the lower limit, the se-

quence can no longer be continued as the rate would become

even less than that when dosing was interrupted.

3.2.2.49. Mode.Parameter.Monitoring.Temp.Status ON,

OFF

Mode.Parameter.Monitoring.Temp.LowLim -170.0...500.0

Mode.Parameter.Monitoring.Temp.UpLim -170.0...500.0

Mode.Parameter.Monitoring.Temp.Action

end, hold, wait, none

Parameters for STAT, DOS, DOC: Monitoring of the limit values of the tem-

perature in °C. Temperature values are entered in the list of measured points

only if the monitoring is active.

On limit value violation, an action can be selected:

end: Abort of the determination.

hold: Interrupt dosing until the sequence is continued by manual switching

(&m $C).

wait: Interrupt dosing until the limit value is no longer violated, then continue

automatically.

3.2.2.50. Mode.Parameter.Monitoring.L10Output.AssignOutput