3.2. Remote control commands

799 GPT Titrino

150

3.2.2.34. Mode.Parameter.TitrPara.ExtrT

0...999 999

Parameter for SET, KFT: Extraction time in s.

3.2.2.35. Mode.Parameter.TitrPara.TDelta

1...2...999 999

Parameter for SET, KFT, STAT: Time interval in s for the entry of a measure-

ment point in the list of measured points.

3.2.2.36. Mode.Parameter.Presel.Cond ON, OFF

Mode.Parameter.Presel.DriftDisp ON, OFF

Mode.Parameter.Presel.DCor.Type auto,

man., OFF

Mode.Parameter.Presel.DCor.Value 0.0...99.9

Parameters for SET, KFT:

.Cond: Conditioning ON/OFF

.DriftDisp: Drift display during conditioning ON/OFF.

.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of

the drift value at start.

.DCor.Value: Drift value for the manual drift correction.


3.2.2.37. Mode.Parameter.CtrlPara.EP pH:

0...±20.00, OFF

U,

Ipol:

0...±2000, OFF

Upol:

0...±200.0, OFF

Mode.Parameter.CtrlPara.UnitEp read

only

Parameters for KFT, STAT: Setting of the EP, resp. Control point as a pH

value, in mV (with U and Ipol) or uA (with Upol). The corresponding unit can

be read with .UnitEP.

With KFT the standard value for Ipol is 250 mV, for Upol 25 uA.

3.2.2.38. Mode.Parameter.CtrlPara.Dyn pH:

0.01...20.00, OFF

U,

Ipol:

1...2000, OFF

Upol:

0.1...200.0, OFF

Mode.Parameter.CtrlPara.UnitDyn read

only

Mode.Parameter.CtrlPara.MaxRate 0.01...10...150, max.

Mode.Parameter.CtrlPara.MinIncr 0.1...9.9, min.

Parameters for KFT, STAT: Control parameters.

.Dyn: Dynamics (control range) in pH, mV (with U and Ipol) or uA (with Upol).

The corresponding unit can be read with .UnitDyn. Standard with

KFT Ipol 100, with KFT Upol 10.

.MaxRate: Maximum allowed titration rate in ml/min. Max. means maximum

possible rate with mounted Exchange Unit.

.MinIncr: Parameter for KFT. Minimum volume increment in ul.

3.2.2.39. Mode.Parameter.Presel.Oven COM1,

COM2,

no

Parameter for KFT.

If an Oven is connected, its results will be incorporated into the result report of

the Titrino.

If there is no Oven connected via RS232, this parameter has to be on "no".