3.1 General rules

758 KFD Titrino

78

E135

Check temp.sensor in MEAS T or with activated temperature monitoring.

Exit: Correct error or &m $S.

E136

Same buffer in CAL. Measured value of the second buffer differs less than

6 mV from the measured value of the first buffer.

Exit: Correct error or &m $S.

E137

XXX Bytes are missing so that the method or the silo line could not be stored

or not enough RAM for running TIP.

Exit: Send new command.

E155

No new silo result (C24 or C25).

Exit: The error message disappears on next start or on recalculation.

E157

No sequence defined in TIP.

Exit. The error message disappears on next start.

E158

A second TIP has been called up in TIP.

Exit: The error message disappears on next start.

E160

No new temporary variable.

Exit: The error message disappears on next start.

E161

Measurement range of the secondary measured value (temperature)

exceeded. The primary measured value (pH, U, Ipol, Upol) can also be

unstable.

Exit: Rectify error or &m $S.

E166

Save lines is "OFF" although a submethod of TIP includes an assignment to

C24 or C25.

Exit: The error message disappears on next start. Attention: The data of this

sample will not be stored.

E172

In TIP a QuickMeas was started, without defining a measuring quantity.

Exit: The error message disappears on next start or &Mode.QuickMeas $S.

E173

The warning interval of the internal buret D0 called.

Exit: Execute prep &a.b.p $G or start next method.

E174

The warning interval of the external buret D1 called.

Exit: Execute prep &a.b.p $G or start next method.

E175

The warning interval of the external buret D2 called.

Exit: Execute prep &a.b.p $G or start next method.

E176

The function &Assembly.Buret.Prep or &Assembly.Buret.Empty was

interrupted manually.

Exit. The error message disappears on next start.

E177

Accessing to the memory card, the card was not (properly) inserted.

Exit: The error message disappears on next start.

E178

The date of changing the battery of the card is expired.

Exit: The error message disappears on next start.

E180

Memory card write-protected.

Exit: Send new command.

E181

Memory card not formatted.

Exit: Send new command.

E182

Memory card not accessible.

Exit: Send new command.

E183

A directory with the same name exists already on the memory card.

Exit: Send new command.

E196

Result is out of limits.

Exit: The error message disappears on next start or on recalculation.

Advertising