3.2 Remote control commands

758 KFD Titrino

101

3.2.2.11. Mode.CFmla

Mode.CFmla.1.Value

0...±999999

Mode.CFmla.2.Value

etc., up to .19

Calculation constants specific to a method. Stored in the method memory of

the Titrino. Operands specific to the sample (3.2.2.93 and 3.2.2.106) and val-

ues of common variables (3.2.2.80) on the other hand are not stored with the

methods.


3.2.2.12. Mode.Parameter.SET1.EP pH:

0...±20.00, OFF

U,

Ipol:

0...±2000, OFF

Upol:

0...±200.0, OFF

Mode.Parameter.SET1.UnitEp read

only

Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. µA (with Upol). The corresponding unit can be read with .UnitEP.

If the value is on "OFF", no further nodes will appear from SET1.

3.2.2.13. Mode.Parameter.SET1.Dyn pH:

0.01...20.00, OFF

U,

Ipol:

1...2000, OFF

Upol:

0.1...200.0, OFF

Mode.Parameter.SET1.UnitDyn read

only

Mode.Parameter.SET1.MaxRate 0.01...10...150, max.

Mode.Parameter.SET1.MinRate 0.01...25.0...999.9

Parameters for SET: Control parameters, see page 33.
.Dyn:

Dynamics, control range in pH, mV (with U and Ipol) or µA (with

Upol). The corresponding unit can be read with .UnitDyn.

.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum

possible rate with the Exchange Unit in current use.

.MinRate: Minimum titration rate in µL/min.

3.2.2.14. Mode.Parameter.SET1.Stop.Type drift, time

Mode.Parameter.SET1.Stop.Drift 1...20...999

Mode.Parameter.SET1.Stop.Time 0...10...999, inf.

Mode.Parameter.SET1.Stop.StopT 0...99999, OFF

Parameters for SET, KFT: Type and size of the stop criterion of the titration.

.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in µL/min. Applies when "drift" has been selected.

.Time: Switch-off delay time in s. Applies when "time" has been selected.

"inf." means infinite.

.StopT: Stop time in s. Applies when "time" has been selected and the value of

.Time is set to "inf.".

3.2.2.15. Mode.Parameter.TitrPara.Direction +,

–, auto

Parameters for SET, KFT: Titration direction.

"auto" means the titration direction is determined automatically by the instru-

ment. If 2 EP's have been set in a SET titration, the titration direction is given

by the two EP's. The entry of the titration direction is then invalid.