Metrohm 758 KFD Titrino Manuel d'utilisation

Page 125

Advertising
background image

3.2 Remote control commands

758 KFD Titrino

121

3.2.2.113. Info.ActualInfo.Titrator.CyclNo read

only

Info.ActualInfo.Titrator.V read

only

Info.ActualInfo.Titrator.Meas read

only

Info.ActualInfo.Titrator.dVdt read

only

Info.ActualInfo.Titrator.dMeasdt read

only

Info.ActualInfo.Titrator.dMeasdV read

only

Info.ActualInfo.Titrator.T read

only

$Q sends the current values in the following formats:

SET

KFT

MEAS

CAL

CyclNo

127 127 127 127

V(mL)

1.2345 1.2345 -

-

Meas:

pH

U, Ipol (mV)
Upol (µA)

T (°C)

3.6(mV)

–345.6

–12.5

-

-

–345.6

–12.5

-

3.345

–345.6

–12.5

25.0

3.345

-

-

-

dVdt (µL/s)

2.5142 2.5142 -

-

dMeasdt

pH,U,Ipol (mV/s)
Upol (µA/s)

T (°C/s)

0.7957

0.7957

-

0.7957

0.7957

-

0.7957

0.7957

0.7957

0.7957

-

-

dMeasdV (mV/µL)

10.6326 10.6326 -

-

T

(°C)

- - 25.9

-

NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal

drift is NV.

OV will be sent for "overrange".

A time frame can be set up from the cycle number and the cycle time (see

3.2.2.118). The cycle number is set to 0 at the start of a method and it is in-

cremented until the end of the method.

3.2.2.114. Info.ActualInfo.MeasPt.Index read

only

Info.ActualInfo.MeasPt.X read

only

Info.ActualInfo.MeasPt.Y read

only

Info.ActualInfo.MeasPt.Z1 read

only

Info.ActualInfo.MeasPt.Z2 read

only

Info.ActualInfo.EP.Index read

only

Info.ActualInfo.EP.X read

only

Info.ActualInfo.EP.Y read

only

$Q sends the last entry into the measuring point list (.MeasPt) or the last entry

into the list of EP's with SET, KFT.

.MeasPt.X"165" Time (SET, KFT, MEAS) of the MPList in s

.MeasPt.Y"3.654" Volume (SET, KFT) of the MPList in mL

.MeasPt.Z1"6.34" Measured value (SET, KFT, MEAS) of the MPList, format

depends on the measured quantity

.MeasPt.Z2"25.8" Temperature of the MPList in °C

.EP.X"1.234" Volume coordinate of the EP

.EP.Y"5.34" Measured value coordinate of the EP

Advertising